Skip to main content

Determination of the δ34S of low-concentration sulfate in water; RSIL lab code 1949

Dates

Year
2006

Citation

Revesz, Kinga, Qi, Haiping, and Coplen, Tyler B, 2006, Determination of the δ34S of low-concentration sulfate in water; RSIL lab code 1949: U.S. Geological Survey.

Summary

The purpose of the Reston Stable Isotope Laboratory (RSIL) lab code 1949 is to determine the δ(34S/32S), abbreviated as δ34S, of dissolved sulfate having a concentration less than 20 milligrams per liter. Dissolved sulfate is collected on an anion-exchange resin in the field, eluted in the laboratory with 3 M KCl, and precipitated with BaCl2 at pH 3 to 4 as BaSO4. The precipitated BaSO4 is filtered and dried before introduction into an elemental analyzer (EA) Carlo Erba NC 2500. The EA is used to convert sulfur in a BaSO4 solid sample into SO2 gas, and the EA is connected to a continuous flow isotope-ratio mass spectrometer (CF-IRMS), which determines differences in the isotope-amount ratios of stable sulfur isotopes (34S/32S) of the [...]

Contacts

Attached Files

Communities

  • USGS National Research Program

Tags

Categories
Types

Provenance

Added to ScienceBase on Fri Apr 19 12:45:35 MDT 2013 by processing file <b>Physical Chemistry of Stable Isotope Fractionation in Hydrologic Processes.xml</b> in item <a href="https://www.sciencebase.gov/catalog/item/504216bae4b04b508bfd337f">https://www.sciencebase.gov/catalog/item/504216bae4b04b508bfd337f</a>

Additional Information

Identifiers

Type Scheme Key
Report Number http://sciencebase.gov/vocab/identifierScheme TM - 10-C8

Citation Extension

citationTypeReport

Item Actions

View Item as ...

Save Item as ...

View Item...