FTS RSS Temperature Test 2, John C. Stennis Space Center, Jun 2016
Dates
Publication Date
2018-01-26
Start Date
2016-06-24
End Date
2016-06-25
Citation
Kunkle, G.A., 2018, Evaluation of the Radar Stage Sensor Manufactured by Forest Technology Systems, Incorporated—Results of Laboratory and Field Testing: U.S. Geological Survey data release, https://doi.org/10.5066/F71C1VSR.
Summary
Two identical Radar Stage Sensors from Forest Technology Systems, were evaluated to determine if they are suitable for U.S. Geological Survey (USGS) hydrologic data collection. The sensors were evaluated in laboratory conditions to evaluate the distance accuracy of the sensor over the manufacturer’s specified operating temperatures and distance to water ranges. Laboratory results were compared to the manufacturer’s accuracy specification of ±0.007 foot (ft) and the USGS Office of Surface Water (OSW) policy requirement that water level sensors have a measurement uncertainty of no more than 0.01 ft or 0.20 percent of the indicated reading. In the temperature chamber test, both sensors were within the manufacturer’s specification for [...]
Summary
Two identical Radar Stage Sensors from Forest Technology Systems, were evaluated to determine if they are suitable for U.S. Geological Survey (USGS) hydrologic data collection. The sensors were evaluated in laboratory conditions to evaluate the distance accuracy of the sensor over the manufacturer’s specified operating temperatures and distance to water ranges. Laboratory results were compared to the manufacturer’s accuracy specification of ±0.007 foot (ft) and the USGS Office of Surface Water (OSW) policy requirement that water level sensors have a measurement uncertainty of no more than 0.01 ft or 0.20 percent of the indicated reading. In the temperature chamber test, both sensors were within the manufacturer’s specification for more than 90 percent of the data points collected over a temperature range of –40 to +60 °C at a fixed distance of 8 ft. This test was performed with radar firmware version 1.09.
Click on title to download individual files attached to this item.
FTS RSS Temperature Test 2, John C. Stennis Space Center, Jun 2016.xml Original FGDC Metadata
View
8.1 KB
application/fgdc+xml
FTSTempData__Test 2_2016.csv
18.88 KB
text/csv
Purpose
Data were obtained as part of an instrument evaluation of the Forest Technology Systems (FTS) Radar Stage Sensor. This data set was used to evaluate whether the sensor met USGS Office of Surface Water accuracy requirements over its full operating temperature range. Instrument and equipment evaluations are done primarily to determine if particular devices would be suitable for use by USGS personnel for hydrologic data collection. Results presented herein were obtained with the instrument model and firmware used at the time of testing, and may or may not represent results obtained from software updates and instrument model improvements. The instrument evaluation results presented are not indicative of an endorsement by the USGS of the tested instrument.